#9450
Strongyloïdes
ratti IgG ELISA
ELISA kit for the diagnosis of human strongyloidosis
Lot 2612S (Expiry date : 26.01.2028)
Production date : 27.05.2026
Results of quality control 1 (after
production), date: 02.06.2026
No serum blank (average) : 0.114
| Sample |
Absorbances of individual wells after blank reduction |
Average |
% pos. |
No serum blank
|
0.000 |
-0.002 |
-0.001 |
0.003 |
0.000 |
0 |
Negative control serum
(N° 9450-06) |
0.062 |
0.056 |
0.065 |
0.058 |
0.060 |
2 |
Weak positive serum
(N° 9450-07) |
0.861 |
0.885 |
0.806 |
0.834 |
0.847 |
35 |
Positive control serum
(N° 9450-08) |
2.440 |
2.558 |
2.384 |
2.372 |
2.439 |
100 |
Validation criteria:
| |
Criterion |
Measure |
Valid |
| Absorbance of positive serum 9450-08 |
> 1.200 |
2.439 |
Yes |
| Absorbance of weak positive serum 9450-07, as % of positive
serum |
> 18 % |
35 % |
Yes |
| Absorbance of negative serum 9450-06, as % of positive serum |
< 10 % |
2 % |
Yes |
| Absorbance of no serum blank |
< 0.350 |
0.114 |
Yes |
Comments:
Good for use.