#9450
Strongyloïdes
ratti IgG ELISA
ELISA kit for the diagnosis of human strongyloidosis
Lot 2611S (Expiry date : 03.01.2028)
Production date : 04.05.2026
Results of quality control 1 (after
production), date: 06.05.2026
No serum blank (average) : 0.128
| Sample |
Absorbances of individual wells after blank reduction |
Average |
% pos. |
No serum blank
|
-0.003 |
0.001 |
0.002 |
0.001 |
0.000 |
0 |
Negative control serum
(N° 9450-06) |
0.076 |
0.081 |
0.085 |
0.079 |
0.080 |
3 |
Weak positive serum
(N° 9450-07) |
0.867 |
0.888 |
0.904 |
0.824 |
0.871 |
36 |
Positive control serum
(N° 9450-08) |
2.440 |
2.362 |
2.431 |
2.476 |
2.427 |
100 |
Validation criteria:
| |
Criterion |
Measure |
Valid |
| Absorbance of positive serum 9450-08 |
> 1.200 |
2.427 |
Yes |
| Absorbance of weak positive serum 9450-07, as % of positive
serum |
> 18 % |
36 % |
Yes |
| Absorbance of negative serum 9450-06, as % of positive serum |
< 10 % |
3 % |
Yes |
| Absorbance of no serum blank |
< 0.350 |
0.128 |
Yes |
Comments:
Good for use.