#9700
Taenia solium IgG ELISA
ELISA kit for the diagnosis of human cysticercosis
Lot 2610C (Expiry date: 03.01.2028)
Production date : 04.05.2026
Results of quality control 1 (after
production), date: 07.05.2026
No serum blank (average) : 0.228
| Sample |
Absorbances of individual wells after blank reduction |
Average |
% pos. |
No serum blank
|
-0.001 |
0.009 |
-0.005 |
-0.003 |
0.000 |
0 |
Negative control serum
(N° 9700-06) |
0.140 |
0.169 |
0.127 |
0.123 |
0.140 |
6 |
Weak positive serum
(N° 9700-07) |
0.498 |
0.522 |
0.468 |
0.414 |
0.476 |
20 |
Positive control
serum
(N° 9700-08) |
2.342 |
2.534 |
2.273 |
2.353 |
2.376 |
100 |
Validation criteria:
| |
Criterion |
Measure |
Valid |
| Absorbance of positive serum 9700-08 |
> 1.200 |
2.376 |
Yes |
|
Absorbance of weak positive serum 9700-07, as % of positive serum
|
> 8 % |
20 % |
Yes
|
| Absorbance of negative serum 9700-06, as % of positive serum |
< 8 % |
6 % |
Yes |
| Absorbance of no serum blank |
< 0.350 |
0.228 |
Yes |
Comments:
Good for use.