#9750
Trichinella spiralis IgG ELISA
ELISA kit for the diagnosis of human trichinellosis
Lot 2608R (Expiry date: 15.12.2027)
Production date : 16.04.2026
Results of quality control 1 (after
production), date: 24.04.2026
No serum blank (average) : 0.135
| Sample |
Absorbances of individual wells after blank reduction |
Average |
% pos. |
No serum blank
|
0.002 |
0.002 |
-0.005 |
0.002 |
0.000 |
0 |
Negative control serum
(N° 9750-06) |
0.076 |
0.079 |
0.057 |
0.075 |
0.071 |
3 |
Weak positive serum
(N° 9750-07) |
0.593 |
0.611 |
0.550 |
0.541 |
0.573 |
24 |
Positive control serum
(N° 9750-08) |
2.397 |
2.633 |
2.239 |
2.401 |
2.417 |
100 |
Validation criteria:
| |
Criterion |
Measure |
Valid |
| Absorbance of positive serum 9750-08 |
> 1.200 |
2.417 |
Yes |
| Absorbance of weak positive serum 9750-07, as % of positive serum |
> 10 % |
24 % |
Yes |
| Absorbance of negative serum 9750-06, as % of positive serum |
< 8 % |
3 % |
Yes |
| Absorbance of no serum blank |
< 0.350 |
0.135 |
Yes |
Comments:
Good for use.
Bordier Affinity Products SA, Chatanerie 2,
Bātiment Biokema,
CH-1023 Crissier, Switzerland
Phone: + 41 21 633 31 67. Fax: +
41 21 633 31 78. Email: cb(at)bordier.ch
|