#9300 Echinococcus multilocularis (Em2-Em18) IgG ELISA
ELISA kit for the diagnosis of human alveolar echinococcosis

Lot 2604M (Expiry date: 08.12.2027)

 

Production date : 09.04.2026

Results of quality control 2 (after 6 months), date: 06.10.2026

No serum blank (average) : 0.139

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
0.004 -0.005 0.005 -0.004 0.000 0
Negative control serum
(N° 9300-06)
0.050 0.054 0.059 0.050 0.053 2
Weak positive serum
(N° 9300-07)
0.706 0.770 0.722 0.694 0.723 33
Positive control serum
(N° 9300-08)
2.153 2.273 2.201 2.104 2.182 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9300-08 > 1.200 2.182 Yes
Absorbance of weak positive serum 9300-07, as % of positive serum > 14 % 33 % Yes
Absorbance of negative serum 9300-06, as % of positive serum < 6 % 2 % Yes
Absorbance of no serum blank < 0.350 0.139 Yes

Comments:

Good for use.