Home    Products    Sales Conditions & distributors    Quality controls    Instructions For Use    Links

#6100 Aspergillus fumigatus IgG ELISA
ELISA kit for the diagnosis of human aspergillosis

Lot 2603A (Expiry date: 08.12.2027)

 

Production date : 09.04.2026

Results of quality control 1 (after production), date: 14.04.2026

No serum blank (average) : 0.140

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
-0.005 0.000 0.005 -0.001 0.000 0
Negative control serum
(N° 61
00-06)
0.047 0.064 0.072 0.069 0.063 3
Weak positive serum
(N° 61
00-07)
0.656 0.658 0.650 0.653 0.655 26
Positive control serum
(N° 61
00-08)
2.469 2.549 2.474 2.491 2.496 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 6100-08 > 1.200 2.496 Yes
Absorbance of weak positive serum 6100-07, as % of positive serum > 11 % 26 % Yes
Absorbance of negative serum 6100-06, as % of positive serum < 8 % 3 % Yes
Absorbance of no serum blank < 0.350 0.140 Yes

Comments:

Good for use.

Bordier Affinity Products SA, Chatanerie 2, Bātiment Biokema, CH-1023 Crissier, Switzerland