#9310
Echinococcus
multilocularis (Em18) IgG ELISA
ELISA kit for human alveolar echinococcosis patient follow-up
Lot 2540F (Expiry date: 02.08.2027)
Production date : 03.12.2025
Results of quality control 2 (after 6
months), date: 01.06.2026
No serum blank (average) : 0.131
| Sample |
Absorbances of individual wells after blank reduction |
Average |
% pos. |
No serum blank
|
-0.002 |
-0.002 |
0.001 |
0.004 |
0.000 |
0 |
Negative control serum
(N° 9310-06) |
0.012 |
0.015 |
0.018 |
0.013 |
0.014 |
1 |
Weak positive serum
(N° 9310-07) |
0.086 |
0.087 |
0.097 |
0.090 |
0.090 |
4 |
Positive control serum
(N° 9310-08) |
2.122 |
2.177 |
2.069 |
2.187 |
2.139 |
100 |
Validation criteria:
| |
Criterion |
Measure |
Valid |
| Absorbance of positive serum 9310-08 |
> 1.200 |
2.139 |
Yes |
| Absorbance of weak positive serum 9310-07, as % of positive serum
|
> 3 %
|
4 % |
Yes |
| Absorbance of negative serum 9310-06, as % of positive serum |
< 3 % |
1 % |
Yes |
| Absorbance of no serum blank |
< 0.350 |
0.131 |
Yes |
Comments:
Good for use.