#9650 Fasciola hepatica IgG ELISA (Trematoda)
ELISA kit for the diagnosis of human fascioliasis

Lot 2538D (Expiry date: 02.08.2027)

 

Production date : 03.12.2025

Results of quality control 2 (after 6 months), date: 01.06.2026

No serum blank (average) : 0.135

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
-0.001 0.000 -0.002 0.003 0.000 0
Negative control serum
(N° 965
0-06)
0.092 0.101 0.074 0.100 0.092 4
Weak positive serum
(N° 965
0-07)
0.426 0.441 0.410 0.410 0.422 18
Positive control serum
(N° 9650
-08)
2.293 2.433 2.336 2.375 2.359 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9650-08 > 1.200 2.359 Yes
Absorbance of weak positive serum 9650-07, as % of positive serum > 8 % 18 % Yes
Absorbance of negative serum 9650-06, as % of positive serum < 8 % 4 % Yes
Absorbance of no serum blank < 0.350 0.135 Yes

Comments:

Good for use.