#9650
Fasciola
hepatica IgG ELISA (Trematoda)
ELISA kit for the diagnosis of human fascioliasis
Lot 2538D (Expiry date: 02.08.2027)
Production date : 03.12.2025
Results of quality control 2 (after
6 months), date: 01.06.2026
No serum blank (average) : 0.135
| Sample |
Absorbances of individual wells after blank reduction |
Average |
% pos. |
No serum blank
|
-0.001 |
0.000 |
-0.002 |
0.003 |
0.000 |
0 |
Negative control serum
(N° 9650-06) |
0.092 |
0.101 |
0.074 |
0.100 |
0.092 |
4 |
Weak positive serum
(N° 9650-07) |
0.426 |
0.441 |
0.410 |
0.410 |
0.422 |
18 |
Positive control
serum
(N° 9650-08) |
2.293 |
2.433 |
2.336 |
2.375 |
2.359 |
100 |
Validation criteria:
| |
Criterion |
Measure |
Valid |
| Absorbance of positive serum 9650-08 |
> 1.200 |
2.359 |
Yes |
| Absorbance of weak positive serum 9650-07, as % of positive serum |
> 8 % |
18 % |
Yes |
| Absorbance of negative serum 9650-06, as % of positive serum |
< 8 % |
4 % |
Yes |
| Absorbance of no serum blank |
< 0.350 |
0.135 |
Yes |
Comments:
Good for use.