Home    Products    Sales Conditions & distributors    Quality controls    Instructions For Use    Links

#9550 Entamoeba histolytica IgG ELISA
ELISA kit for the diagnosis of human visceral amebiasis

Lot 2536H (Expiry date: 26.07.2027)

 

Production date : 27.11.2025

Results of quality control 2 (after 6 months), date: 26.05.2026

No serum blank (average) : 0.141

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
-0.011 0.005 0.001 0.006 0.000 0
Negative control serum
(N° 9550-06)
0.120 0.128 0.134 0.137 0.129 6
Weak positive serum
(N° 9550-07)
0.763 0.846 0.818 0.829 0.814 35
Positive control serum
(N° 9550-08)
2.279 2.258 2.306 2.344 2.296 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9550-08 > 1.200 2.296 Yes
Absorbance of weak positive serum 9550-07, as % of positive serum >  17 % 35 % Yes
Absorbance of negative serum 9550-06, as % of positive serum < 12 % 6 % Yes
Absorbance of no serum blank < 0.350 0.141 Yes

Comments:

Good for use.

Bordier Affinity Products SA, Chatanerie 2, Bātiment Biokema, CH-1023 Crissier, Switzerland
Phone: + 41 21 633 31 67.  Fax: + 41 21 633 31 78. Email: cb(at)bordier.ch