#9350
Echinococcus
granulosus IgG ELISA
ELISA kit for the diagnosis of human echinococcoses
Lot 2524G (Expiry date: 07.03.2027)
Production date : 08.07.2025
Results of quality control 1 (after
production), date: 15.07.2025
No serum blank (average) : 0.136
Sample |
Absorbances of individual wells after blank reduction |
Average |
% pos. |
No serum blank
|
-0.004 |
0.000 |
0.002 |
0.002 |
0.000 |
0 |
Negative control serum
(N° 9350-06) |
0.102 |
0.106 |
0.111 |
0.117 |
0.109 |
5 |
Weak positive serum
(N° 9350-07) |
1.088 |
1.076 |
1.119 |
1.132 |
1.104 |
54 |
Positive control serum
(N° 9350-08) |
2.014 |
2.049 |
2.036 |
2.141 |
2.060 |
100 |
Validation criteria:
|
Criterion |
Measure |
Valid |
Absorbance of positive serum 9350-08 |
> 1.200 |
2.060 |
Yes |
Absorbance of weak positive serum 9350-07, as % of positive serum |
> 27 % |
54 % |
Yes |
Absorbance of negative serum 9350-06, as % of positive serum |
< 8 % |
5 % |
Yes |
Absorbance of no serum blank |
< 0.350 |
0.136 |
Yes |
Comments:
Good for use.