#9400 Acanthocheilonema viteae (Onchocercidae) IgG ELISA
ELISA kit for the diagnosis of human filariasis

Lot 2523V (Expiry date: 07.03.2027)

 

Production date : 08.07.2025

Results of quality control 1 (after production), date: 11.07.2025

No serum blank (average) : 0.133

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
0.001 0.000 -0.005 0.004 0.000 0
Negative control serum
(N° 9400-06)
0.103 0.090 0.088 0.089 0.093 4
Weak positive serum
(N° 9400-07)
0.905 0.895 0.911 0.884 0.899 40
Positive control serum
(N° 9400-08)
2.212 2.346 2.161 2.330 2.262 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9400-08 > 1.200 2.262 Yes
Absorbance of weak positive serum 9400-07, as % of positive serum > 19 % 40 % Yes
Absorbance of negative serum 9400-06, as % of positive serum < 10 % 4 % Yes
Absorbance of no serum blank < 0.350 0.133 Yes

Comments:

Good for use.