#9200
Toxocara
canis IgG ELISA
ELISA kit for the diagnosis of human toxocarosis
Lot 2521T (Expiry date: 24.02.2027)
Production date : 25.06.2025
Results of quality control 3 (after
12 months), date: 19.06.2026
No serum blank (average) : 0.146
| Sample |
Absorbances of individual wells after blank reduction |
Average |
% pos. |
No serum blank
|
-0.005 |
-0.001 |
0.004 |
0.002 |
0.000 |
0 |
Negative control serum
(N° 9200-06) |
0.011 |
0.025 |
0.032 |
0.029 |
0.024 |
1 |
Weak positive serum
(N° 9200-07) |
1.734 |
1.754 |
1.723 |
1.620 |
1.708 |
69 |
Positive control serum
(N° 9200-08) |
2.423 |
2.576 |
2.427 |
2.503 |
2.482 |
100 |
Validation criteria:
| |
Criterion |
Measure |
Valid |
| Absorbance of positive serum 9200-08 |
> 1.200 |
2.482 |
Yes |
| Absorbance of weak positive serum 9200-07, as % of positive serum |
> 32 % |
69 % |
Yes |
| Absorbance of negative serum 9200-06, as % of positive serum |
< 6 % |
1 % |
Yes |
| Absorbance of no serum blank |
< 0.350 |
0.146 |
Yes |
Comments:
Good for use.