Home    Products    Sales Conditions & distributors    Quality controls    Instructions For Use    Links

#9200 Toxocara canis IgG ELISA
ELISA kit for the diagnosis of human toxocarosis

Lot 2521T (Expiry date: 24.02.2027)

 

Production date : 25.06.2025

Results of quality control 1 (after production), date: 02.07.2025

No serum blank (average) : 0.133

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
-0.001 0.004 -0.002 0.000 0.000 0
Negative control serum
(N° 9200-06)
0.039 0.036 0.026 0.040 0.035 1
Weak positive serum
(N° 9200-07)
1.735 1.655 1.705 1.595 1.672 70
Positive control serum
(N° 9200-08)
2.333 2.506 2.328 2.331 2.374 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9200-08 > 1.200 2.374 Yes
Absorbance of weak positive serum 9200-07, as % of positive serum >  32 % 70 % Yes
Absorbance of negative serum 9200-06, as % of positive serum < 6 % 1 % Yes
Absorbance of no serum blank < 0.350 0.133 Yes

Comments:

Good for use.

Bordier Affinity Products SA, Chatanerie 2, Bātiment Biokema, CH-1023 Crissier, Switzerland