#9550 Entamoeba histolytica
IgG ELISA
ELISA kit for the diagnosis of human visceral amebiasis
Lot 2520H (Expiry date: 24.02.2027)
Production date : 25.06.2025
Results of quality control 3 (after
12 months), date: 19.06.2026
No serum blank (average) : 0.150
| Sample |
Absorbances of individual wells after blank reduction |
Average |
% pos. |
No serum blank
|
-0.006 |
0.000 |
0.006 |
0.001 |
0.000 |
0 |
Negative control serum
(N° 9550-06) |
0.098 |
0.117 |
0.114 |
0.112 |
0.110 |
5 |
Weak positive serum
(N° 9550-07) |
0.688 |
0.688 |
0.700 |
0.737 |
0.703 |
32 |
Positive control serum
(N° 9550-08) |
2.168 |
2.195 |
2.154 |
2.235 |
2.188 |
100 |
Validation criteria:
| |
Criterion |
Measure |
Valid |
| Absorbance of positive serum 9550-08 |
> 1.200 |
2.188 |
Yes |
| Absorbance of weak positive serum 9550-07, as % of positive serum |
> 17 % |
32 % |
Yes |
| Absorbance of negative serum 9550-06, as % of positive serum |
< 12 % |
5 % |
Yes |
| Absorbance of no serum blank |
< 0.350 |
0.150 |
Yes |
Comments:
Good for use.