Home    Products    Sales Conditions & distributors    Quality controls    Instructions For Use    Links

#9550 Entamoeba histolytica IgG ELISA
ELISA kit for the diagnosis of human visceral amebiasis

Lot 2520H (Expiry date: 24.02.2027)

 

Production date : 25.06.2025

Results of quality control 1 (after production), date: 30.06.2025

No serum blank (average) : 0.136

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
-0.003 -0.004 0.008 0.000 0.000 0
Negative control serum
(N° 9550-06)
0.096 0.090 0.114 0.108 0.102 5
Weak positive serum
(N° 9550-07)
0.737 0.733 0.754 0.764 0.747 35
Positive control serum
(N° 9550-08)
2.076 2.138 2.115 2.201 2.132 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9550-08 > 1.200 2.132 Yes
Absorbance of weak positive serum 9550-07, as % of positive serum >  17 % 35 % Yes
Absorbance of negative serum 9550-06, as % of positive serum < 12 % 5 % Yes
Absorbance of no serum blank < 0.350 0.136 Yes

Comments:

Good for use.

Bordier Affinity Products SA, Chatanerie 2, Bātiment Biokema, CH-1023 Crissier, Switzerland
Phone: + 41 21 633 31 67.  Fax: + 41 21 633 31 78. Email: cb(at)bordier.ch