Home    Products    Sales Conditions & distributors    Quality controls    Instructions For Use    Links

#9750 Trichinella spiralis IgG ELISA
ELISA kit for the diagnosis of human trichinellosis

Lot 2518R (Expiry date: 09.02.2027)

 

Production date : 10.06.2025

Results of quality control 1 (after production), date: 16.06.2025

No serum blank (average) : 0.135

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
-0.004 -0.001 0.004 0.001 0.000 0
Negative control serum
(N° 9750-06)
0.041 0.054 0.041 0.049 0.046 2
Weak positive serum
(N° 9750-07)
0.404 0.414 0.438 0.470 0.432 18
Positive control serum
(N° 9750-08)
2.465 2.393 2.409 2.551 2.455 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9750-08 > 1.200 2.455 Yes
Absorbance of weak positive serum 9750-07, as % of positive serum >  10 % 18 % Yes
Absorbance of negative serum 9750-06, as % of positive serum < 8 % 2 % Yes
Absorbance of no serum blank < 0.350 0.135 Yes

Comments:

Good for use.

Bordier Affinity Products SA, Chatanerie 2, Bātiment Biokema, CH-1023 Crissier, Switzerland
Phone: + 41 21 633 31 67.  Fax: + 41 21 633 31 78. Email: cb(at)bordier.ch