#9450
Strongyloïdes
ratti IgG ELISA
ELISA kit for the diagnosis of human strongyloidosis
Lot 2517S (Expiry date : 03.02.2027)
Production date : 04.06.2025
Results of quality control 1 (after
production), date: 06.06.2025
No serum blank (average) : 0.134
Sample |
Absorbances of individual wells after blank reduction |
Average |
% pos. |
No serum blank
|
-0.001 |
0.008 |
-0.006 |
-0.002 |
0.000 |
0 |
Negative control serum
(N° 9450-06) |
0.084 |
0.097 |
0.089 |
0.104 |
0.094 |
4 |
Weak positive serum
(N° 9450-07) |
0.861 |
0.853 |
0.950 |
0.930 |
0.899 |
38 |
Positive control serum
(N° 9450-08) |
2.324 |
2.419 |
2.261 |
2.484 |
2.372 |
100 |
Validation criteria:
|
Criterion |
Measure |
Valid |
Absorbance of positive serum 9450-08 |
> 1.200 |
2.372 |
Yes |
Absorbance of weak positive serum 9450-07, as % of positive
serum |
> 18 % |
38 % |
Yes |
Absorbance of negative serum 9450-06, as % of positive serum |
< 10 % |
4 % |
Yes |
Absorbance of no serum blank |
< 0.350 |
0.134 |
Yes |
Comments:
Good for use.