#9600 Schistosoma mansoni IgG ELISA
ELISA kit for the diagnosis of human schistosomiasis

Lot 2513B (Expiry date: 26.11.2026)

 

Production date : 27.03.2025

Results of quality control 2 (after 6 months), date: 23.09.2025

No serum blank (average) : 0.159

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
0.006 -0.001 -0.003 -0.003 0.000 0
Negative control serum
(N° 9600-06)
0.056 0.042 0.052 0.046 0.049 2
Weak positive serum
(N° 9600-07)
0.397 0.393 0.382 0.378 0.388 18
Positive control serum
(N° 9600-08)
2.162 2.210 2.134 2.216 2.181 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9600-08 > 1.200 2.181 Yes
Absorbance of weak positive serum 9600-07, as % of positive serum > 10 % 18 % Yes
Absorbance of negative serum 9600-06, as % of positive serum < 8 % 2 % Yes
Absorbance of no serum blank < 0.350 0.159 Yes

Comments:

Good for use.