#6100 Aspergillus fumigatus IgG ELISA
ELISA kit for the diagnosis of human aspergillosis

Lot 2509A (Expiry date: 19.10.2026)

 

Production date : 20.02.2025

Results of quality control 2 (after 6 months), date: 19.08.2025

No serum blank (average) : 0.131

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
-0.001 0.004 -0.005 0.002 0.000 0
Negative control serum
(N° 61
00-06)
0.044 0.059 0.052 0.053 0.052 2
Weak positive serum
(N° 61
00-07)
0.529 0.590 0.569 0.592 0.570 26
Positive control serum
(N° 61
00-08)
2.114 2.321 2.200 2.285 2.230 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 6100-08 > 1.200 2.230 Yes
Absorbance of weak positive serum 6100-07, as % of positive serum > 11 % 26 % Yes
Absorbance of negative serum 6100-06, as % of positive serum < 8 % 2 % Yes
Absorbance of no serum blank < 0.350 0.131 Yes

Comments:

Good for use.