#9450
Strongyloïdes
ratti IgG ELISA
ELISA kit for the diagnosis of human strongyloidosis
Lot 2508S (Expiry date : 19.10.2026)
Production date : 20.02.2025
Results of quality control 2 (after
6 months), date: 19.08.2025
No serum blank (average) : 0.131
Sample |
Absorbances of individual wells after blank reduction |
Average |
% pos. |
No serum blank
|
-0.007 |
-0.003 |
0.005 |
0.005 |
0.000 |
0 |
Negative control serum
(N° 9450-06) |
0.070 |
0.071 |
0.088 |
0.080 |
0.077 |
3 |
Weak positive serum
(N° 9450-07) |
0.696 |
0.679 |
0.695 |
0.721 |
0.698 |
32 |
Positive control serum
(N° 9450-08) |
2.103 |
2.250 |
2.124 |
2.354 |
2.208 |
100 |
Validation criteria:
|
Criterion |
Measure |
Valid |
Absorbance of positive serum 9450-08 |
> 1.200 |
2.208 |
Yes |
Absorbance of weak positive serum 9450-07, as % of positive
serum |
> 18 % |
32 % |
Yes |
Absorbance of negative serum 9450-06, as % of positive serum |
< 10 % |
3 % |
Yes |
Absorbance of no serum blank |
< 0.350 |
0.131 |
Yes |
Comments:
Good for use.
Bordier Affinity Products SA, Chatanerie 2,
Bâtiment Biokema,
CH-1023 Crissier, Switzerland