Home    Products    Sales Conditions & distributors    Quality controls    Instructions For Use    Links

#9700 Taenia solium IgG ELISA
ELISA kit for the diagnosis of human cysticercosis

Lot 2507C (Expiry date: 21.09.2026)

Production date : 22.01.2025

Results of quality control 2 (after 6 months), date: 21.07.2025

No serum blank (average) : 0.214

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
0.002 -0.002 0.006 -0.005 0.000 0
Negative control serum
(N° 9700-06)
0.116 0.116 0.122 0.131 0.121 5
Weak positive serum
(N° 9700-07)
0.469 0.486 0.478 0.472 0.476 18
Positive control serum
(N° 9700-08)
2.587 2.635 2.556 2.725 2.626 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9700-08 > 1.200 2.626 Yes
Absorbance of weak positive serum 9700-07, as % of positive serum > 8 % 18 % Yes
Absorbance of negative serum 9700-06, as % of positive serum < 8 % 5 % Yes
Absorbance of no serum blank < 0.350 0.214 Yes

Comments:

Good for use.

Bordier Affinity Products SA, Chatanerie 2, Bātiment Biokema, CH-1023 Crissier, Switzerland
Phone: + 41 21 633 31 67.  Fax: + 41 21 633 31 78. Email: cb(at)bordier.ch