#9750
Trichinella spiralis IgG ELISA
ELISA kit for the diagnosis of human trichinellosis
Lot 2505R (Expiry date: 14.09.2026)
Production date : 15.01.2025
Results of quality control 4 (after
20 months), date: 16.09.2026
No serum blank (average) : 0.132
| Sample |
Absorbances of individual wells after blank reduction |
Average |
% pos. |
No serum blank
|
0.002 |
-0.005 |
0.003 |
0.000 |
0.000 |
0 |
Negative control serum
(N° 9750-06) |
0.041 |
0.042 |
0.048 |
0.041 |
0.043 |
2 |
Weak positive serum
(N° 9750-07) |
0.351 |
0.373 |
0.324 |
0.370 |
0.355 |
17 |
Positive control serum
(N° 9750-08) |
1.962 |
2.269 |
2.004 |
2.083 |
2.080 |
100 |
Validation criteria:
| |
Criterion |
Measure |
Valid |
| Absorbance of positive serum 9750-08 |
> 1.200 |
2.080 |
Yes |
| Absorbance of weak positive serum 9750-07, as % of positive serum |
> 10 % |
17 % |
Yes |
| Absorbance of negative serum 9750-06, as % of positive serum |
< 8 % |
2 % |
Yes |
| Absorbance of no serum blank |
< 0.350 |
0.132 |
Yes |
Comments:
Good for use.