#9450
Strongyloïdes
ratti IgG ELISA
ELISA kit for the diagnosis of human strongyloidosis
Lot 2504S (Expiry date : 14.09.2026)
Production date : 15.01.2025
Results of quality control 2 (after
6 months), date: 15.07.2025
No serum blank (average) : 0.131
Sample |
Absorbances of individual wells after blank reduction |
Average |
% pos. |
No serum blank
|
-0.004 |
0.003 |
0.001 |
0.001 |
0.000 |
0 |
Negative control serum
(N° 9450-06) |
0.080 |
0.079 |
0.060 |
0.085 |
0.076 |
3 |
Weak positive serum
(N° 9450-07) |
0.713 |
0.717 |
0.711 |
0.730 |
0.717 |
32 |
Positive control serum
(N° 9450-08) |
2.142 |
2.362 |
2.154 |
2.336 |
2.248 |
100 |
Validation criteria:
|
Criterion |
Measure |
Valid |
Absorbance of positive serum 9450-08 |
> 1.200 |
2.248 |
Yes |
Absorbance of weak positive serum 9450-07, as % of positive
serum |
> 18 % |
32 % |
Yes |
Absorbance of negative serum 9450-06, as % of positive serum |
< 10 % |
3 % |
Yes |
Absorbance of no serum blank |
< 0.350 |
0.131 |
Yes |
Comments:
Good for use.