#9350 Echinococcus granulosus IgG ELISA
ELISA kit for the diagnosis of human echinococcoses

Lot 2503G (Expiry date: 07.09.2026)

 

Production date : 08.01.2025

Results of quality control 2 (after 6 months), date: 08.07.2025

No serum blank (average) : 0.131

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
-0.002 -0.002 0.003 0.000 0.000 0
Negative control serum
(N° 9350-06)
0.098 0.093 0.116 0.095 0.101 5
Weak positive serum
(N° 9350-07)
1.087 1.156 1.124 1.195 1.141 53
Positive control serum
(N° 9350-08)
2.069 2.205 2.142 2.236 2.163 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9350-08 > 1.200 2.163 Yes
Absorbance of weak positive serum 9350-07, as % of positive serum >  27 % 53 % Yes
Absorbance of negative serum 9350-06, as % of positive serum < 8 % 5 % Yes
Absorbance of no serum blank < 0.350 0.131 Yes

Comments:

Good for use.