#9400 Acanthocheilonema viteae (Onchocercidae) IgG ELISA
ELISA kit for the diagnosis of human filariasis

Lot 2502V (Expiry date: 07.09.2026)

 

Production date : 08.01.2025

Results of quality control 2 (after 6 monhts), date: 08.07.2025

No serum blank (average) : 0.129

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
-0.006 -0.003 0.001 0.008 0.000 0
Negative control serum
(N° 9400-06)
0.102 0.093 0.107 0.107 0.102 5
Weak positive serum
(N° 9400-07)
0.854 0.887 0.870 0.864 0.869 40
Positive control serum
(N° 9400-08)
2.078 2.211 2.170 2.248 2.177 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9400-08 > 1.200 2.177 Yes
Absorbance of weak positive serum 9400-07, as % of positive serum > 19 % 40 % Yes
Absorbance of negative serum 9400-06, as % of positive serum < 10 % 5 % Yes
Absorbance of no serum blank < 0.350 0.129 Yes

Comments:

Good for use.