#9600 Schistosoma mansoni IgG ELISA
ELISA kit for the diagnosis of human schistosomiasis

Lot 2439B (Expiry date: 12.07.2026)

 

Production date : 13.11.2024

Results of quality control 4 (after 20 months), date: 13.07.2026

No serum blank (average) : 0.157

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
-0.002 0.002 -0.001 0.001 0.000 0
Negative control serum
(N° 9600-06)
0.043 0.039 0.064 0.052 0.049 2
Weak positive serum
(N° 9600-07)
0.365 0.365 0.384 0.376 0.372 17
Positive control serum
(N° 9600-08)
2.143 2.073 2.169 2.177 2.140 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9600-08 > 1.200 2.140 Yes
Absorbance of weak positive serum 9600-07, as % of positive serum > 10 % 17 % Yes
Absorbance of negative serum 9600-06, as % of positive serum < 8 % 2 % Yes
Absorbance of no serum blank < 0.350 0.157 Yes

Comments:

Good for use.