#9600 Schistosoma mansoni IgG ELISA
ELISA kit for the diagnosis of human schistosomiasis

Lot 2439B (Expiry date: 12.07.2026)

 

Production date : 13.11.2024

Results of quality control 2 (after 6 months), date: 12.05.2025

No serum blank (average) : 0.165

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
-0.019 -0.008 -0.004 0.031 0.000 0
Negative control serum
(N° 9600-06)
0.051 0.044 0.060 0.041 0.049 2
Weak positive serum
(N° 9600-07)
0.409 0.399 0.429 0.425 0.416 19
Positive control serum
(N° 9600-08)
2.189 2.237 2.180 2.304 2.228 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9600-08 > 1.200 2.228 Yes
Absorbance of weak positive serum 9600-07, as % of positive serum > 10 % 19 % Yes
Absorbance of negative serum 9600-06, as % of positive serum < 8 % 2 % Yes
Absorbance of no serum blank < 0.350 0.165 Yes

Comments:

Good for use.