#9300 Echinococcus multilocularis (Em2-Em18) IgG ELISA
ELISA kit for the diagnosis of human alveolar echinococcosis

Lot 2433M (Expiry date: 08.06.2026)

 

Production date : 09.10.2024

Results of quality control 4 (after 20 months), date: 09.06.2026

No serum blank (average) : 0.141

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
-0.007 0.002 -0.002 0.007 0.000 0
Negative control serum
(N° 9300-06)
0.012 0.014 0.037 0.072 0.034 2
Weak positive serum
(N° 9300-07)
0.582 0.554 0.584 0.584 0.576 27
Positive control serum
(N° 9300-08)
2.023 2.122 2.123 2.219 2.122 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9300-08 > 1.200 2.122 Yes
Absorbance of weak positive serum 9300-07, as % of positive serum > 14 % 27 % Yes
Absorbance of negative serum 9300-06, as % of positive serum < 6 % 2 % Yes
Absorbance of no serum blank < 0.350 0.141 Yes

Comments:

Good for use.

Bordier Affinity Products SA, Chatanerie 2, Bātiment Biokema, CH-1023 Crissier, Switzerland