#9300 Echinococcus multilocularis (Em2-Em18) IgG ELISA
ELISA kit for the diagnosis of human alveolar echinococcosis

Lot 2433M (Expiry date: 08.06.2026)

 

Production date : 09.10.2024

Results of quality control 3 (after 12 months), date: 03.10.2025

No serum blank (average) : 0.130

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
-0.005 0.005 -0.001 0.002 0.000 0
Negative control serum
(N° 9300-06)
0.018 0.021 0.019 0.026 0.021 1
Weak positive serum
(N° 9300-07)
0.596 0.578 0.610 0.557 0.585 26
Positive control serum
(N° 9300-08)
2.238 2.228 2.203 2.200 2.217 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9300-08 > 1.200 2.217 Yes
Absorbance of weak positive serum 9300-07, as % of positive serum > 14 % 26 % Yes
Absorbance of negative serum 9300-06, as % of positive serum < 6 % 1 % Yes
Absorbance of no serum blank < 0.350 0.130 Yes

Comments:

Good for use.