Home    Products    Sales Conditions & distributors    Quality controls    Instructions For Use    Links

#9700 Taenia solium IgG ELISA
ELISA kit for the diagnosis of human cysticercosis

Lot 2427C-R (Expiry date: 21.03.2026)

Production date : 22.07.2024

Results of quality control 3 (after 12 months & rework), date: 18.07.2025

No serum blank (average) : 0.194

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
0.006 -0.003 -0.006 0.004 0.000 0
Negative control serum
(N° 9700-06)
0.115 0.114 0.092 0.097 0.104 5
Weak positive serum
(N° 9700-07)
0.395 0.389 0.351 0.336 0.368 17
Positive control serum
(N° 9700-08)
2.198 2.230 2.229 2.188 2.211 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9700-08 > 1.200 2.211 Yes
Absorbance of weak positive serum 9700-07, as % of positive serum > 8 % 17 % Yes
Absorbance of negative serum 9700-06, as % of positive serum < 8 % 5 % Yes
Absorbance of no serum blank < 0.350 0.194 Yes

Comments:

Good for use.

Bordier Affinity Products SA, Chatanerie 2, Bātiment Biokema, CH-1023 Crissier, Switzerland
Phone: + 41 21 633 31 67.  Fax: + 41 21 633 31 78. Email: cb(at)bordier.ch