#9200
Toxocara
canis IgG ELISA
ELISA kit for the diagnosis of human toxocarosis
Lot 2426T-R (Expiry date: 21.03.2026)
Production date : 22.07.2024
Results of quality control 3 (after
12 months & rework), date: 18.07.2025
No serum blank (average) : 0.134
Sample |
Absorbances of individual wells after blank reduction |
Average |
% pos. |
No serum blank
|
-0.003 |
-0.001 |
0.003 |
0.000 |
0.000 |
0 |
Negative control serum
(N° 9200-06) |
0.036 |
0.033 |
0.024 |
0.021 |
0.029 |
1 |
Weak positive serum
(N° 9200-07) |
1.293 |
1.253 |
1.337 |
1337 |
1.305 |
61 |
Positive control serum
(N° 9200-08) |
2.084 |
2.152 |
2.099 |
2.259 |
2.149 |
100 |
Validation criteria:
|
Criterion |
Measure |
Valid |
Absorbance of positive serum 9200-08 |
> 1.200 |
2.149 |
Yes |
Absorbance of weak positive serum 9200-07, as % of positive serum |
> 32 % |
61 % |
Yes |
Absorbance of negative serum 9200-06, as % of positive serum |
< 6 % |
1 % |
Yes |
Absorbance of no serum blank |
< 0.350 |
0.134 |
Yes |
Comments:
Good for use.