#9400 Acanthocheilonema viteae (Onchocercidae) IgG ELISA
ELISA kit for the diagnosis of human filariasis

Lot 2424V (Expiry date: 07.03.2026)

 

Production date : 08.07.2024

Results of quality control 3 (after 12 months), date: 04.07.2025

No serum blank (average) : 0.121

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
-0.002 -0.005 0.006 0.001 0.000 0
Negative control serum
(N° 9400-06)
0.110 0.085 0.113 0.090 0.099 4
Weak positive serum
(N° 9400-07)
0.880 0.859 0.849 0.847 0.858 38
Positive control serum
(N° 9400-08)
2.221 2.334 2.175 2.370 2.275 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9400-08 > 1.200 2.275 Yes
Absorbance of weak positive serum 9400-07, as % of positive serum > 19 % 38 % Yes
Absorbance of negative serum 9400-06, as % of positive serum < 10 % 4 % Yes
Absorbance of no serum blank < 0.350 0.121 Yes

Comments:

Good for use.