Home    Products    Sales Conditions & distributors    Quality controls    Instructions For Use    Links

#9310 Echinococcus multilocularis (Em18) IgG ELISA
ELISA kit for human alveolar echinococcosis patient follow-up

Lot 2420F-R (Expiry date: 02.02.2026)

Production date : 03.06.2024

Results of quality control 3 (after 12 months & rework), date: 03.06.2025

No serum blank (average) : 0.110

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
0.000 -0.002 0.000 0.002 0.000 0
Negative control serum
(N° 9310-06)
0.023 0.035 0.020 0.026 0.026 1
Weak positive serum
(N° 9310-07)
0.086 0.111 0.086 0.098 0.095 4
Positive control serum
(N° 9310-08)
2.062 2.255 2.061 2.158 2.134 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9310-08 > 1.200 2.134 Yes
Absorbance of weak positive serum 9310-07, as % of positive serum > 3 % 4 % Yes
Absorbance of negative serum 9310-06, as % of positive serum < 3 % 1 % Yes
Absorbance of no serum blank < 0.350 0.110 Yes

Comments:

Good for use.

Bordier Affinity Products SA, Chatanerie 2, Bātiment Biokema, CH-1023 Crissier, Switzerland