Home    Products    Sales Conditions & distributors    Quality controls    Instructions For Use    Links

#9200 Toxocara canis IgG ELISA
ELISA kit for the diagnosis of human toxocarosis

Lot 2401T (Expiry date: 07.09.2025)

 

Production date : 08.01.2024

Results of quality control 4 (after 20 months), date: 08.09.2025

No serum blank (average) : 0.128

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
-0.009 0.014 -0.005 0.001 0.000 0
Negative control serum
(N° 9200-06)
0.002 0.006 0.019 0.024 0.013 1
Weak positive serum
(N° 9200-07)
1.410 1.394 1.442 1.292 1.384 58
Positive control serum
(N° 9200-08)
2.318 2.467 2.355 2.405 2.386 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9200-08 > 1.200 2.386 Yes
Absorbance of weak positive serum 9200-07, as % of positive serum >  32 % 58 % Yes
Absorbance of negative serum 9200-06, as % of positive serum < 6 % 1 % Yes
Absorbance of no serum blank < 0.350 0.128 Yes

Comments:

Good for use.

Bordier Affinity Products SA, Chatanerie 2, Bātiment Biokema, CH-1023 Crissier, Switzerland