#9750
Trichinella spiralis IgG ELISA
ELISA kit for the diagnosis of human trichinellosis
Lot 2337R (Expiry date: 26.07.2025)
Production date : 27.11.2023
Results of quality control 4 (after
20 months), date: 28.07.2025
No serum blank (average) : 0.124
Sample |
Absorbances of individual wells after blank reduction |
Average |
% pos. |
No serum blank
|
0.003 |
-0.002 |
0.002 |
-0.002 |
0.000 |
0 |
Negative control serum
(N° 9750-06) |
0.037 |
0.039 |
0.040 |
0.041 |
0.039 |
2 |
Weak positive serum
(N° 9750-07) |
0.400 |
0.428 |
0.408 |
0.371 |
0.402 |
17 |
Positive control serum
(N° 9750-08) |
2.186 |
2.499 |
2.295 |
2.274 |
2.313 |
100 |
Validation criteria:
|
Criterion |
Measure |
Valid |
Absorbance of positive serum 9750-08 |
> 1.200 |
2.313 |
Yes |
Absorbance of weak positive serum 9750-07, as % of positive serum |
> 10 % |
17 % |
Yes |
Absorbance of negative serum 9750-06, as % of positive serum |
< 8 % |
2 % |
Yes |
Absorbance of no serum blank |
< 0.350 |
0.124 |
Yes |
Comments:
Good for use.
Bordier Affinity Products SA, Chatanerie 2,
Bātiment Biokema,
CH-1023 Crissier, Switzerland
Phone: + 41 21 633 31 67. Fax: +
41 21 633 31 78. Email: cb(at)bordier.ch
|