#9650 Fasciola hepatica IgG ELISA (Trematoda)
ELISA kit for the diagnosis of human fascioliasis

Lot 2334D (Expiry date: 29.06.2025)

 

Production date : 30.10.2023

Results of quality control 4 (after 20 months), date: 30.06.2025

No serum blank (average) : 0.123

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
0.004 -0.002 -0.005 0.003 0.000 0
Negative control serum
(N° 965
0-06)
0.095 0.097 0.088 0.100 0.095 4
Weak positive serum
(N° 965
0-07)
0.414 0.416 0.3410 0.402 0.411 17
Positive control serum
(N° 9650
-08)
2.446 2.584 2.333 2.541 2.476 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9650-08 > 1.200 2.476 Yes
Absorbance of weak positive serum 9650-07, as % of positive serum > 8 % 17 % Yes
Absorbance of negative serum 9650-06, as % of positive serum < 8 % 4 % Yes
Absorbance of no serum blank < 0.350 0.123 Yes

Comments:

Good for use.