Home    Products    Sales Conditions & distributors    Quality controls    Instructions For Use    Links

#9310 Echinococcus multilocularis (Em18) IgG ELISA
ELISA kit for human alveolar echinococcosis patient follow-up

Lot 2326F (Expiry date: 03.05.2025)

Production date : 04.09.2023

Results of quality control 2 (after 6 months), date: 28.02.2024

No serum blank (average) : 0.135

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
-0.018 0.014 -0.012 0.016 0.000 0
Negative control serum
(N° 9310-06)
-0.003 0.073 0.007 0.071 0.037 2
Weak positive serum
(N° 9310-07)
0.118 0.157 0.117 0.163 0.138 6
Positive control serum
(N° 9310-08)
2.094 2.251 2.242 2.444 2.257 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9310-08 > 1.200 2.257 Yes
Absorbance of weak positive serum 9310-07, as % of positive serum > 3 % 6 % Yes
Absorbance of negative serum 9310-06, as % of positive serum < 3 % 2 % Yes
Absorbance of no serum blank < 0.350 0.135 Yes

Comments:

Good for use.

Bordier Affinity Products SA, Chatanerie 2, Bātiment Biokema, CH-1023 Crissier, Switzerland
Phone: + 41 21 633 31 67.  Fax: + 41 21 633 31 78. Email: cb(at)bordier.ch