#9200 Toxocara canis IgG ELISA
ELISA kit for the diagnosis of human toxocarosis

Lot 2323T (Expiry date: 12.03.2025)

 

Production date : 13.07.2023

Results of quality control 1 (after production), date: 19.07.2023

No serum blank (average) : 0.114

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
-0.001 -0.003 0.004 -0.001 0.000 0
Negative control serum
(N° 9200-06)
0.032 0.047 0.027 0.048 0.038 2
Weak positive serum
(N° 9200-07)
1.431 1.388 1.531 1.520 1.467 64
Positive control serum
(N° 9200-08)
2.143 2.303 2.239 2.422 2.276 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9200-08 > 1.200 2.276 Yes
Absorbance of weak positive serum 9200-07, as % of positive serum >  32 % 64 % Yes
Absorbance of negative serum 9200-06, as % of positive serum < 6 % 2 % Yes
Absorbance of no serum blank < 0.350 0.114 Yes

Comments:

Good for use.

Bordier Affinity Products SA, Chatanerie 2, Bātiment Biokema, CH-1023 Crissier, Switzerland