#9600 Schistosoma mansoni IgG ELISA
ELISA kit for the diagnosis of human schistosomiasis

Lot 2321B (Expiry date: 05.03.2025)

 

Production date : 06.07.2023

Results of quality control 2 (after 6 months), date: 05.01.2024

No serum blank (average) : 0.149

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
-0.010 0.021 -0.007 -0.003 0.000 0
Negative control serum
(N° 9600-06)
0.074 0.137 0.069 0.097 0.094 4
Weak positive serum
(N° 9600-07)
0.515 0.484 0.455 0.453 0.477 21
Positive control serum
(N° 9600-08)
2.145 2.346 2.179 2.263 2.233 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9600-08 > 1.200 2.233 Yes
Absorbance of weak positive serum 9600-07, as % of positive serum > 10 % 21 % Yes
Absorbance of negative serum 9600-06, as % of positive serum < 8 % 4 % Yes
Absorbance of no serum blank < 0.350 0.149 Yes

Comments:

Good for use.