#9450 Strongyloïdes ratti IgG ELISA
ELISA kit for the diagnosis of human strongyloidosis

Lot 2316S (Expiry date : 31.12.2024)

Production date : 01.05.2023

Results of quality control 2 (after 6 months), date: 27.10.2023

No serum blank (average) : 0.116

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
-0.005 -0.003 0.002 0.006 0.000 0
Negative control serum
(N° 9450-06)
0.092 0.112 0.110 0.137 0.112 5
Weak positive serum
(N° 9450-07)
0.813 0.876 0.870 0.885 0.861 36
Positive control serum
(N° 9450-08)
2.211 2.450 2.336 2.565 2.390 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9450-08 > 1.200 2.390 Yes
Absorbance of weak positive serum 9450-07, as % of positive serum > 18 % 36 % Yes
Absorbance of negative serum 9450-06, as % of positive serum < 10 % 5 % Yes
Absorbance of no serum blank < 0.350 0.116 Yes

Comments:

Good for use.