#9450
Strongyloïdes
ratti IgG ELISA
ELISA kit for the diagnosis of human strongyloidosis
Lot 2316S (Expiry date : 31.12.2024)
Production date : 01.05.2023
Results of quality control 1 (after
production), date: 08.05.2023
No serum blank (average) : 0.109
Sample |
Absorbances of individual wells after blank reduction |
Average |
% pos. |
No serum blank
|
0.001 |
-0.002 |
0.002 |
0.000 |
0.000 |
0 |
Negative control serum
(N° 9450-06) |
0.090 |
0.098 |
0.078 |
0.100 |
0.091 |
4 |
Weak positive serum
(N° 9450-07) |
0.817 |
0.869 |
0.849 |
0.869 |
0.851 |
36 |
Positive control serum
(N° 9450-08) |
2.240 |
2.404 |
2.258 |
2.465 |
2.342 |
100 |
Validation criteria:
|
Criterion |
Measure |
Valid |
Absorbance of positive serum 9450-08 |
> 1.200 |
2.342 |
Yes |
Absorbance of weak positive serum 9450-07, as % of positive
serum |
> 18 % |
36 % |
Yes |
Absorbance of negative serum 9450-06, as % of positive serum |
< 10 % |
4 % |
Yes |
Absorbance of no serum blank |
< 0.350 |
0.109 |
Yes |
Comments:
Good for use.