#9350 Echinococcus granulosus IgG ELISA
ELISA kit for the diagnosis of human echinococcoses

Lot 2312G (Expiry date: 15.11.2024)

 

Production date : 16.03.2023

Results of quality control 3 (after 12 months), date: 11.03.2024

No serum blank (average) : 0.134

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
-0.011 0.015 -0.010 0.007 0.000 0
Negative control serum
(N° 9350-06)
0.041 0.103 0.036 0.083 0.066 3
Weak positive serum
(N° 9350-07)
1.194 1.116 1.039 1.127 1.119 52
Positive control serum
(N° 9350-08)
2.047 2.322 2.076 2.207 2.163 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9350-08 > 1.200 2.163 Yes
Absorbance of weak positive serum 9350-07, as % of positive serum >  27 % 52 % Yes
Absorbance of negative serum 9350-06, as % of positive serum < 8 % 3 % Yes
Absorbance of no serum blank < 0.350 0.134 Yes

Comments:

Good for use.