#9450
Strongyloïdes
ratti IgG ELISA
ELISA kit for the diagnosis of human strongyloidosis
Lot 2310S (Expiry date : 08.11.2024)
Production date : 09.03.2023
Results of quality control 3 (after
12 months), date: 29.02.2024
No serum blank (average) : 0.132
Sample |
Absorbances of individual wells after blank reduction |
Average |
% pos. |
No serum blank
|
-0.016 |
0.015 |
-0.012 |
0.013 |
0.000 |
0 |
Negative control serum
(N° 9450-06) |
0.068 |
0.127 |
0.092 |
0.130 |
0.104 |
5 |
Weak positive serum
(N° 9450-07) |
0.784 |
0.877 |
0.794 |
0.858 |
0.828 |
38 |
Positive control serum
(N° 9450-08) |
2.178 |
2.241 |
2.162 |
2.207 |
2.197 |
100 |
Validation criteria:
|
Criterion |
Measure |
Valid |
Absorbance of positive serum 9450-08 |
> 1.200 |
2.197 |
Yes |
Absorbance of weak positive serum 9450-07, as % of positive
serum |
> 18 % |
38 % |
Yes |
Absorbance of negative serum 9450-06, as % of positive serum |
< 10 % |
5 % |
Yes |
Absorbance of no serum blank |
< 0.350 |
0.132 |
Yes |
Comments:
Good for use.