#9450
Strongyloïdes
ratti IgG ELISA
ELISA kit for the diagnosis of human strongyloidosis
Lot 2310S (Expiry date : 08.11.2024)
Production date : 09.03.2023
Results of quality control 2 (after
6 months), date: 06.09.2023
No serum blank (average) : 0.115
Sample |
Absorbances of individual wells after blank reduction |
Average |
% pos. |
No serum blank
|
0.001 |
-0.003 |
0.004 |
-0.001 |
0.000 |
0 |
Negative control serum
(N° 9450-06) |
0.095 |
0.094 |
0.099 |
0.110 |
0.099 |
4 |
Weak positive serum
(N° 9450-07) |
0.773 |
0.820 |
0.776 |
0.781 |
0.787 |
35 |
Positive control serum
(N° 9450-08) |
2.152 |
2.268 |
2.245 |
2.319 |
2.246 |
100 |
Validation criteria:
|
Criterion |
Measure |
Valid |
Absorbance of positive serum 9450-08 |
> 1.200 |
2.246 |
Yes |
Absorbance of weak positive serum 9450-07, as % of positive
serum |
> 18 % |
35 % |
Yes |
Absorbance of negative serum 9450-06, as % of positive serum |
< 10 % |
4 % |
Yes |
Absorbance of no serum blank |
< 0.350 |
0.115 |
Yes |
Comments:
Good for use.