#9450
Strongyloïdes
ratti IgG ELISA
ELISA kit for the diagnosis of human strongyloidosis
Lot 2310S (Expiry date : 08.11.2024)
Production date : 09.03.2023
Results of quality control 1 (after
production), date: 14.03.2023
No serum blank (average) : 0.118
Sample |
Absorbances of individual wells after blank reduction |
Average |
% pos. |
No serum blank
|
0.001 |
-0.003 |
-0.002 |
0.004 |
0.000 |
0 |
Negative control serum
(N° 9450-06) |
0.114 |
0.102 |
0.098 |
0.115 |
0.107 |
5 |
Weak positive serum
(N° 9450-07) |
0.852 |
0.834 |
0.796 |
0.809 |
0.823 |
36 |
Positive control serum
(N° 9450-08) |
2.293 |
2.332 |
2.203 |
2.334 |
2.291 |
100 |
Validation criteria:
|
Criterion |
Measure |
Valid |
Absorbance of positive serum 9450-08 |
> 1.200 |
2.291 |
Yes |
Absorbance of weak positive serum 9450-07, as % of positive
serum |
> 18 % |
36 % |
Yes |
Absorbance of negative serum 9450-06, as % of positive serum |
< 10 % |
5 % |
Yes |
Absorbance of no serum blank |
< 0.350 |
0.118 |
Yes |
Comments:
Good for use.
Bordier Affinity Products SA, Chatanerie 2,
Bâtiment Biokema,
CH-1023 Crissier, Switzerland
Phone: + 41 21 633 31 67. Fax: +
41 21 633 31 78. Email: cb(at)bordier.ch
|