#9300 Echinococcus multilocularis (Em2-Em18) IgG ELISA
ELISA kit for the diagnosis of human alveolar echinococcosis

Lot 2307M (Expiry date: 14.10.2024)

 

Production date : 15.02.2023

Results of quality control 2 (after 6 months), date: 14.08.2023

No serum blank (average) : 0.131

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
-0.002 -0.003 -0.001 0.005 0.000 0
Negative control serum
(N° 9300-06)
0.026 0.040 0.035 0.034 0.034 1
Weak positive serum
(N° 9300-07)
0.754 0.772 0.739 0.755 0.755 29
Positive control serum
(N° 9300-08)
2.624 2.732 2.550 2.644 2.638 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9300-08 > 1.200 2.638 Yes
Absorbance of weak positive serum 9300-07, as % of positive serum > 14 % 29 % Yes
Absorbance of negative serum 9300-06, as % of positive serum < 6 % 1  % Yes
Absorbance of no serum blank < 0.350 0.131 Yes

Comments:

Good for use.