#9350
Echinococcus
granulosus IgG ELISA
ELISA kit for the diagnosis of echinococcoses in humans
Lot 2222G
Results of quality control
Kit: Bordier Affinity Products SA, Echinococcus granulosus IgG ELISA, Article 9350
Expiry date: 11.03.2024, Control date: 12.03.2024,
Control # 4
No serum blank (average) : 0.126
Sample |
Absorbances of individual wells after blank reduction |
Average |
% pos. |
No serum blank
|
-0.016 |
0.012 |
-0.008 |
0.012 |
0.000 |
0 |
Negative control serum
(N° 9350-06) |
0.012 |
0.068 |
0.018 |
0.074 |
0.043 |
2 |
Weak positive serum
(N° 9350-07) |
0.708 |
0.818 |
0.768 |
0.835 |
0.782 |
45 |
Positive control serum
(N° 9350-08) |
1.622 |
1.884 |
1.708 |
1.768 |
1.745 |
100 |
Validation criteria:
|
Criterion |
Measure |
Valid |
Absorbance of positive serum 9350-08 |
> 1.200 |
1.745 |
Yes |
Absorbance of negative serum 9350-06, as % of positive serum |
< 12 % |
2 % |
Yes |
Absorbance of no serum blank |
< 0.350 |
0.126 |
Yes |
Comments:
Good for use.