#9750
Trichinella spiralis IgG ELISA
ELISA kit for the diagnosis of trichinellosis in humans
Lot 2220R
Results of quality control
Kit: Bordier Affinity Products SA, Trichinella spiralis IgG ELISA, Article
9750 .
Expiry date: 11.03.2024, Control date: 12.03.2024,
Control # 4
No serum blank (average) : 0.142
Sample |
Absorbances of individual wells after blank reduction |
Average |
% pos. |
No serum blank
|
-0.025 |
0.032 |
-0.020 |
0.012 |
0.000 |
0 |
Negative control serum
(N° 9750-06) |
0.019 |
0.112 |
0.014 |
0.076 |
0.056 |
3 |
Weak positive serum
(N° 9750-07) |
0.397 |
0.426 |
0.379 |
0.423 |
0.407 |
19 |
Positive control serum
(N° 9750-08) |
2.049 |
2.249 |
2.064 |
2.271 |
2.159 |
100 |
Validation criteria:
|
Criterion |
Measure |
Valid |
Absorbance of positive serum 9750-08 |
> 1.200 |
2.159 |
Yes |
Absorbance of negative serum 9750-06, as % of positive serum |
< 10 % |
3 % |
Yes |
Absorbance of no serum blank |
< 0.350 |
0.142 |
Yes |
Comments:
Good for use.
Bordier Affinity Products SA, Chatanerie 2,
Bātiment Biokema,
CH-1023 Crissier, Switzerland
Phone: + 41 21 633 31 67. Fax: +
41 21 633 31 78. Email: cb(at)bordier.ch
|