Lot 2212H

 

Results of quality control

Kit: Bordier Affinity Products SA, Entamoeba histolytica IgG ELISA, Article 9550.
Expiry date: 10.12.2023, Control date: 12.12.2023,  Control # 4

No serum blank (average) : 0.157

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
-0.012 0.009 -0.016 0.019 0.000 0
Negative control serum
(N° 9550-06)
0.106 0.134 0.082 0.159 0.120 6
Weak positive serum
(N° 9550-07)
0.396 0.425 0.504 0.439 0.441 22
Positive control serum
(N° 9550-08)
1.826 1.838 2.260 2.044 1.992 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9550-08 > 1.200 1.992 Yes
Absorbance of negative serum 9550-06, as % of positive serum < 14 % 6 % Yes
Absorbance of no serum blank < 0.350 0.157 Yes

Comments:

Good for use.

Bordier Affinity Products SA, Chatanerie 2, Bātiment Biokema, CH-1023 Crissier, Switzerland