#9550 Entamoeba histolytica
IgG ELISA
ELISA kit for the diagnosis of human visceral amebiasis
Lot 2212H
Results of quality control
Kit: Bordier Affinity Products SA, Entamoeba histolytica IgG ELISA, Article
9550 .
Expiry date: 10.12.2023, Control date: 05.04.2023,
Control # 3
No serum blank (average) : 0.146
Sample |
Absorbances of individual wells after blank reduction |
Average |
% pos. |
No serum blank
|
-0.001 |
-0.007 |
-0.001 |
0.010 |
0.000 |
0 |
Negative control serum
(N° 9550-06) |
0.094 |
0.087 |
0.113 |
0.096 |
0.097 |
4 |
Weak positive serum
(N° 9550-07) |
0.440 |
0.453 |
0.493 |
0.450 |
0.459 |
21 |
Positive control serum
(N° 9550-08) |
2.157 |
2.228 |
2.179 |
2.115 |
2.170 |
100 |
Validation criteria:
|
Criterion |
Measure |
Valid |
Absorbance of positive serum 9550-08 |
> 1.200 |
2.170 |
Yes |
Absorbance of negative serum 9550-06, as % of positive serum |
< 14 % |
4 % |
Yes |
Absorbance of no serum blank |
< 0.350 |
0.146 |
Yes |
Comments:
Good for use.
Bordier Affinity Products SA, Chatanerie 2,
Bātiment Biokema,
CH-1023 Crissier, Switzerland
Phone: + 41 21 633 31 67. Fax: +
41 21 633 31 78. Email: cb(at)bordier.ch
|