Lot 2212H

 

Results of quality control

Kit: Bordier Affinity Products SA, Entamoeba histolytica IgG ELISA, Article 9550.
Expiry date: 10.12.2023, Control date: 05.04.2023,  Control #  3

No serum blank (average) : 0.146

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
-0.001 -0.007 -0.001 0.010 0.000 0
Negative control serum
(N° 9550-06)
0.094 0.087 0.113 0.096 0.097 4
Weak positive serum
(N° 9550-07)
0.440 0.453 0.493 0.450 0.459 21
Positive control serum
(N° 9550-08)
2.157 2.228 2.179 2.115 2.170 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9550-08 > 1.200 2.170 Yes
Absorbance of negative serum 9550-06, as % of positive serum < 14 % 4 % Yes
Absorbance of no serum blank < 0.350 0.146 Yes

Comments:

Good for use.